Detail NR1 ELEC/INST RM/KL/CM
2025-04-24
L63-2K2
Wahyu Adi Triono
Problem: Repare Analizer
Action:
Replace key pard analyzer and sensor O2. Test calibration zero and span O2 result ok ( RIZAL & ANDRI & EVE IQBAL )

Status: done